High-Speed Optical Calibration
Rapidly inspect dispensing conditions with non-contact optical measurement while reducing the waiting and accumulation time required by conventional gravimetric measurement.
NeoScale high-speed optical dispensing metrology replaces traditional collection and gravimetric weighing with non-contact image analysis. It enables rapid evaluation of single-shot dispensing volume and jetting consistency, substantially reducing inspection time while delivering digitized data for advanced equipment and process analytics.
Optic
Measurement principleNeoScale converts optical information into quantitative dispensing data, revealing individual-shot and short-cycle variation that conventional averaged weighing can overlook.
A camera, precision illumination, droplet edge detection, image processing, and calibration algorithms identify droplet contours, dimensions, and geometric features from optical images.
A calibration model translates optical geometry into dispensing data, making it possible to evaluate consistency as valve condition, nozzle condition, and process parameters change.
Measurement workflow
A continuous four-stage workflow turns droplet imagery into process information.
Rapid
Response advantageNeoScale shortens the path from parameter adjustment to validated dispensing conditions, particularly for automated qualification involving large nozzle arrays, multi-valve systems, or multiple process recipes.
Conventional method
Typically requires accumulation of multiple droplets to produce a sufficient mass signal, making it better suited to offline or sampling-based inspection.
NeoScale method
Quickly compares dispense quantity changes caused by pressure, pulse duration, valve opening settings, temperature, or material conditions.
Process impact
Faster feedback accelerates automated qualification and process-window development while exposing variations between individual shots or short dispensing cycles.
Link
Equipment integrationNeoScale is designed around measurement module integration and automated verification. Its non-contact architecture can be integrated into automated dispensing equipment without the weighing pan, liquid collection vessel, and stringent vibration isolation required by conventional precision balances.
Equipment can automatically move the jet valve to the measurement position before production, after material changeover, following a recipe change, during scheduled maintenance, or in response to an anomaly.
No balance settling time is required.
Minimal influence from liquid evaporation.
Minimal influence from environmental vibration.
Weight, droplet behavior, and nozzle condition in one workflow.
Verification Triggers — Before production, after material changeover, following a recipe change, during scheduled maintenance, or in response to an anomaly.
Integration Benefits — Measurement does not rely on collecting liquid on a weighing pan. Liquid capture and waste handling must still be addressed in the equipment design.
View
Seven functionsTwo optical perspectives connect nozzle condition with complete droplet geometry and flight behavior.
Dual-View Observation: 0° X/Z and 90° Y/Z
Nozzle Monitoring
Droplet Calculations
Time
Cycle comparisonDispensing: 1 s + Settling: 2 s
Total Cycle Time: 3.0 sDispensing: 0.2 s + Analysis: 0.1 s
Total Cycle Time: 0.3 sThe workflow illustration specifies a balance settling time of 2 s, whereas the table below lists 5 s. Both values are retained from the source material. The manufacturer should confirm whether different test conditions apply.
Versus
Measurement comparison| Characteristic | Conventional Precision Balance | NeoScale |
|---|---|---|
| Precision | 1 µg | 0.001 µg |
| Required Shots | > 100 shots | > 5 shots |
| Effect of Liquid Evaporation | Severe | Almost None |
| Effect of Environmental Vibration | Sensitive | Almost None |
| Result Output | Overall Weight |
|
| Settling Time | 5 s | 0 s |
| Start-up Time | 60 min | 1 min |
Comparison data are based on MOL internal testing and are provided for technical reference only. Actual performance depends on material, equipment configuration, and test conditions. “Precision” is retained from the source terminology and should not be interpreted as a guaranteed accuracy specification.
Control
PROCESS MANAGEMENT APPLICATIONSProcess Traceability & Preventive Maintenance
In process management, NeoScale does more than replace an electronic balance. More importantly, it digitizes dispensing measurement. Each measurement result can be associated with the Valve ID, Nozzle ID, recipe, material, time stamp, and equipment parameters, moving dispensing quality management from conventional manual sampling toward automated, traceable, data-driven control.
Apply
KEY APPLICATIONSNeoScale can use brief periods of process dead time to perform high-speed optical calibration. Its compact mechanical design supports next-to-line deployment, allowing measurement and verification without substantially disrupting production cadence.
Rapidly inspect dispensing conditions with non-contact optical measurement while reducing the waiting and accumulation time required by conventional gravimetric measurement.
Perform checks during material changeover, line change, recipe switching, maintenance, or anomaly response to make effective use of brief process dead time.
The compact mechanical design supports next-to-line deployment and integration with automated handling, valve positioning, and equipment-control workflows.
DATA ASSOCIATION
MANAGEMENT APPLICATIONS
PROCESS IMPACT
Move from manual sampling to automated, data-driven control. NeoScale transforms every dispensing measurement into traceable process data to support statistical control, nozzle-to-nozzle comparison, valve-aging monitoring, and preventive maintenance.